80
Inspiration platform pressure
Range: 0~100 cmH
2
O, Resolution: 1 cmH
2
O
Minimum airway pressure
Range: 0~100 cmH
2
O, Resolution: 1 cmH
2
O
Oxygen concentration
Range: 21~100%, Resolution: 1%
Airway resistance
Range: 0~200 cmH
2
O/(L/s), Resolution: 1 cmH
2
O/(L/s)
Compliance
Range: 0~200 mL/cmH
2
O, Resolution: 1 mL/cmH
2
O
Rapid shallow breathing
Range: 0.0-1000.0 bpm/L, Resolution: 1 bpm/L
Occlusion pressure P0.1
Range: 30>=P0.1>= 0, Resolution: 0.1 cmH
2
O
Negative inspiration force
Range: -40~0cmH2O, Resolution: 0.1 cmH
2
O
AutoPEEP
Range: 0~100cmH2O, Resolution: 0.1 cmH
2
O
PR
Range: 30~250 bpm, Resolution: 1bpm
EtCO2
Range: 0~99 mmHg, Resolution: 1mmHg
InCO2
Range: 0~99 mmHg, Resolution: 1mmHg
B.5 Main alarm parameters
Alarm content Range
Tidal volume low
High
100~2000 mL, OFF
Low
OFF, 20~2000 mL
Minute volume
High
0~40 L
Low
OFF, 0~39 L
O
2
concentration
High
21~100%
Low
OFF, 21~99%
Airway pressure
High
6~100 cmH
2
O
Low
0~99cmH
2
O
Frequency
High
1~100bpm
Low
0~99 bpm
EtCO
2
(optional)
High
1~99 mmHg
Low
OFF, 0~98 mmHg
SPO
2
(optional) Low
80~100%
O
2
supply pressure
High
110~750 KPa
Low
100~740 KPa
Air supply pressure
High
110~750 KPa
Low
100~740 KPa
upper limits
Plimit≤Ppeak+5cmH
2
O