Guidance and manufacturer’s declaration electromagnetic immunity
Boaray 5000 C/D series is intended for use in the electromagnetic environment specified
below. The customer or the user of the
Boaray 5000 C/D series should assure that it is used
in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV
air
±8 kV
contact
±15 kV
air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least
30 %.
Electrostatic
transient / burst
IEC 61000-4-4
±2kV
for power
supply lines
±2kV
for power
supply lines
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV
for input/output
lines
± 2 kV for
lines to
ground
±1 kV for
input/output lines
±
2 kV for
lines to
ground
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power supply
< 5 % U
T (>95 % dip in
U
T ) for 0,5 cycle
40 % U
T (60 % dip in UT )
for 5 cycles
70 % U
T (30 % dip in UT )
for 25 cycles
< 5 % U
T (>95 % dip in
< 5 % U
T (>95 % dip in
U
T ) for 0,5 cycle
40 % U
T (60 % dip in
U
T ) for 5 cycles
70 % U
T (30 % dip in
U
T ) for 25 cycles
< 5 % U
T (>95 % dip in
Mains power quality
should be that of a
typical commercial or
hospital environment.