Remote Control Commands
R&S
®
NRPxxS(N)
121User Manual 1177.5079.02 ─ 10
6. Create a backup of the calibration data set before making any changes.
Select "File" > "Save Calibration Data".
A dialog opens where you can select the location to save the calibration data.
To change the S-parameter data
1. In the device table, double-click an entry. See also "Device table" on page 119.
2. Select "File" > "Import S2P".
3. Select the *.S2P file you want to import, and confirm with "Open".
The data from the selected file is loaded in the device table.
All uncertainties are set to zero.