Remote Control Basics
R&S
®
NRPxxS(N)
168User Manual 1177.5079.02 ─ 10
11.3.6 Questionable Status Register
Contains information on questionable sensor states. Such states occur when the sen-
sor is not operated in compliance with its specifications.
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
+
Questionable
Status
0
0
0
Power Summary
0
0
0
0
POST Failure
0
0
0
0
0
0
Calibration Summary
Querying the register:
●
STATus:QUEStionable:CONDition?
●
STATus:QUEStionable[:EVENt]?
Table 11-6: Meaning of bits used in the questionable status register
Bit No. Meaning
0 to 2 Not used
3 Questionable power status register summary bit
Corresponds to the summary bit of the questionable power status register.
See Chapter 11.3.6.1, "Questionable Power Status Register", on page 169.
4 to 7 Not used
8 Questionable calibration status register summary bit
Corresponds to the summary bit of the questionable calibration status register.
See Chapter 11.3.6.2, "Questionable Calibration Status Register", on page 169.
9 POST failure
The built-in test of the R&S NRPxxS(N) carried out automatically upon power-up has
generated an error.
10 to 15 Not used
Status Reporting System