Instructions for Use MPR1 DATALOGGER
Manufacturer: zwo-399EN
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany Rev. 9, 2020-10-16
Page 54 of 98
Technical alarm
In comparison with cabled ICP measurement the following differences exist:
With ICPT there is no display of the pulse amplitude of the intracranial pressure.
The resolution of the ICPT display is 1 mmHg, without any decimal places.
As from firmware version 2.00.0063 respectively 2.00.0063 HW2, the ID of the telemetric catheter is
displayed in addition to ICPT after the telemetric connection has been established.
Cabled ICP measurement
Telemetric ICPT measurement
If no telemetric connection can be created or if the telemetric connection is interrupted or lost, the following
display is output: