Instructions for Use MPR1 DATALOGGER
Manufacturer: zwo-399EN
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany Rev. 9, 2020-10-16
Page 68 of 98
4.9.2 Factory settings
4.9.2.1 Factory settings and adjustment ranges for physiological alarms
The factory setting of the display language is English.
The table rendered below shows the factory settings for physiological alarms as well as their adjustment
ranges.
Parameter
Measuring
location
Unit
Factory settings
Adjustment range
Alarm limit
Alarm limits
*1
Adjustment
step size
lower
upper
activated
lower
upper
Pressure
ART
SYS
mmHg
100
180
ja
-40
400
5
Pressure
ART
MAP
mmHg
80
140
ja
-40
400
5
Pressure
ART
DIA
mmHg
60
100
ja
-40
400
5
Pressure
ZVD
mmHg
0
20
ja
-40
400
1
Pressure
ICP1
ICP
mmHg
0
30
ja
-40
400
1
Pressure
ICP1
ICPA
mmHg
0
20
nein
0
400
1
Pressure
ICP2
ICP
mmHg
0
30
ja
-40
400
1
Pressure
ICP2
ICPA
mmHg
0
20
nein
0
400
1
Pressure
ICP1/2
CPP
mmHg
50
100
ja
0
400
1
Pressure
D1
SYS
mmHg
100
180
ja
-40
400
5
Pressure
D1
MAP
mmHg
80
140
ja
-40
400
5
Pressure
D1
DIA
mmHg
60
100
ja
-40
400
5
Pressure
D2
SYS
mmHg
100
180
ja
-40
400
5
Pressure
D2
MAP
mmHg
80
140
ja
-40
400
5
Pressure
D2
DIA
mmHg
60
120
ja
-40
400
5
Pressure
D3
SYS
mmHg
100
180
ja
-40
400
5
Pressure
D3
MAP
mmHg
80
140
ja
-40
400
5
Pressure
D3
DIA
mmHg
60
120
ja
-40
400
5
Pressure
ICPT
ICP
mmHg
0
30
ja
-40
400
1
Temperature
T1
°C
35,8
37,5
ja
15,0
45,0
0,1
Temperature
T2
°C
35,8
37,5
ja
15,0
45,0
0,1
Table 6: Factory settings
*1 The alarm limits can be adjusted freely within the range given; the minimum distance between upper
and lower limit has to be one step width.
4.9.2.2 Inactivate alarm signal switch
Global audio suppression for a duration of 10 minutes