10.1.3 Electromagnetic emissions
The MPR1 is intended for use in an electromagnetic environment as specified below. The operator or the
user of the MPR1 should ensure that it is operated in such an environment.
Measurement of Immunity Immunity test level
CISPR 11, group 1, class A
CISPR 11, group 1, class A
Harmonic current emissions
Voltage changes, voltage fluctuations
and flicker emissions
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.
10.1.4 Electromagnetic immunity
Measurement of Immunity Immunity test level
Electrostatic discharge immunity
according IEC 61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Radiated RF electromagnetic field
immunity according IEC 61000-4-3
80-2700 MHz; 1 kHz AM 80%; 3 V/m at 2 Hz
Immunity to proximity fields from RF
wireless communications equipment
according IEC 61000-4-3
385 MHz; Pulse modulation: 18 Hz; 27 V/m
450 MHz, FM + 5Hz deviation: 1 kHz sine; 28 V/m
710, 745, 780 MHz; Pulse modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse modulation: 217 Hz; 28 V/m
2450 MHz; Pulse modulation: 217 Hz; 28 V/m
5240, 5500, 5785 MHz; Pulse modulation: 217 Hz; 9 V/m
Electrical fast transient/burst immunity
according – a. c. mains according
±2 kV; 100 kHz repetition frequency
Electrical fast transient/burst immunity
according – I/O SIP/SOP ports according
±2 kV; 100 kHz repetition frequency
Surge immunity according IEC 61000-4-5
Line-to-line: ±0.5, ±1 kV
Immunity to conducted disturbances
induced by RF fields – a. c. mains
0,15 - 80 MHz; 1kHz AM 80%; 3 Vrms , 6 Vrms in ISM
bands between 0,15 MHz and 80 MHz at 2 Hz
Immunity to conducted disturbances
induced by RF fields – SIP/SOP ports
0,15 - 80 MHz; 1kHz AM 80%; 3 Vrms , 6 Vrms in ISM
bands between 0,15 MHz and 80 MHz at 2 Hz
Power frequency magnetic field
immunity nach IEC 61000-4-8
Voltage dips, short interruptions
and voltage variations immunity nach
IEC 61000-4-11
T
for 0,5 cycle at 0, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % U
T
for 1 Cycle at 0°
70 % U
T
for 25/30 cycles bei 0°
0 % U
T
for 250/300 cycles bei 0°