What to do if my RAUMEDIC MPR1 Data Loggers telemetry catheter is not recognised?
T
Terri HortonAug 7, 2025
If your RAUMEDIC Data Loggers telemetry catheter isn't being recognized, there are several potential causes and solutions:
* The reader might be the issue. Try checking or replacing the reader.
* The firmware version MPR1 could be too low. Verify the firmware version MPR1.
* The distance between the reader and the telemetric catheter may be too great. Reduce the distance.
* The TK itself might be defective. In this case, replace the TK.
K
Kenneth OsborneAug 17, 2025
What to do if my RAUMEDIC Data Loggers reader is not recognised?
A
Anne FloresAug 17, 2025
If the reader for your RAUMEDIC Data Loggers isn't being recognized, consider the following:
* Ensure the reader is correctly connected to the appropriate socket. If it wasn't properly connected, connect the reader to the correct socket.
* The reader itself might be defective, so try changing the reader.
* The firmware version MPR1 could be outdated. Check the firmware version MPR1.
R
Rhonda MarshallNov 25, 2025
What to do if the functional test of the sterile RAUMEDIC MPR1 Data Loggers TK fails before implantation?
K
Kevin StanleyNov 25, 2025
If the functional test required of the sterile TK fails before implantation, consider these potential causes and solutions:
* A major deviation from the expected value of 0 mm Hg could be due to drift of the pressure sensor of the TK. Replace the TK.
* A major deviation from the expected value of 0 mm Hg could also be caused by drift of the pressure sensor of the reader. Check the reader, for example, by checking the air pressure measured by the reader, which can be displayed by screen 4, and by using a comparator system for comparison or by using another reader as replacement.
* The TK or reader might be defective. Replace the TK and/or reader.
* A major deviation from the expected value of 0 mm Hg could be due to drift of the pressure sensor of the reader, STK of the reader not ...