Instructions for Use MPR1 DATALOGGER
Manufacturer: zwo-399EN
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany Rev. 9, 2020-10-16
Page 80 of 98
Contents of messages which are displayed after
pressing the Info key
Additional recommendation
The implant software version is not compatible with the
measurement system. Use another pressure
measurement implant or contact the service staff.
Get in touch with the Service
Department.
There is an inconsistency in the reader data. A
measurement process cannot be carried out. Use
The reader software version is not compatible with the
measurement system. A measurement process cannot
be carried out. Use another reader or contact the
service staff for a software update.
Get in touch with the Service
Department.
No telemetric connection possible. Please check the
positioning of the antenna.
Internal error in the measurement ICPT. The
measuring accuracy for the ICPT is not guaranteed.
Use another RFID reader. If the error occurs
repeatedly, there is an error in the device itself. Please
contact the manufacturer.
Get in touch with the Service
Department.
Internal error RFID reader. The function of the RFID
reader can no longer be guaranteed. If the error occurs
repeatedly, there is an error in the device itself. Please
contact the manufacturer.
Get in touch with the Service
Department.
The data memory is almost completely full. Save the
data on the PC, otherwise the oldest data will soon be
Messages on the MPR1
• Messages on the MPR1 when setting up the telemetric connection.
If a TDT1 readP reader is connected to the MPR1, the unit will try to build up a connection to the telemetric
catheter. At the same time, a timer will be started, and the time for searching a telemetric catheter is
displayed. If no telemetric connection is built up within 30 s, the fault message T33 is displayed, and a
single reminder signal is sounded for a period of 1 s. The fault message T33 displayed can be closed by
pressing OK. The purpose of this message is to call attention of the user to the unsuccessful connection
setup to a telemetric catheter (see 4.4.5 Telemetric pressure measurement
)