4.6.2 Device diagnostics
In order to diagnose the apparatus, use the Forward key to skip to the screen Device Setup. After
having press the Diag F key, the Device diagnostics is accessed.
Here the following information is rendered:
Software version
Power supply OK
You can start a test of the temperature channels by pressing the button. This is carried out automatically
for the pressure channels.
Press Diag F key to access
Device diagnostics
T-T
Software version
F key to start the test of the
temperature channels
F key to start the
alarm system check
F key to start the test of the
ecg channel
Power supply check
additional informations ICPT
Software ver sion
The current status of the MPR1 firmware is software version 2.10.0070 or 2.10.0070 HW2.
The version of the software with the hardware level HW2 is intended for devices without ECG applied part.
Devices of earlier delivery date can have a lower version of the software. With respect to the firmware, the
MPR1 can be updated. Please get in touch with the manufacturer to check the availability of a new firmware
version.