54
7
Repeat the same procedure for the
lancing device (step
1
to step
6
).
NOTE
After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the meter
works properly before using the meter.
Control solution tests should be performed using two
different levels of ReliOn™ Premier Control Solutions (Level 1
and/or Level 2).
Verify that the test results are within the range printed on
the test strip vial.
• See page 28-30 for how to perform a control solution test.
NOTE
If any of the following deterioration signs appear after
pre-cleaning or disinfecting, please stop using the system and
contact Customer Service: 1-855-776-0662, 7 a.m. to 7 p.m.
Central time, 7 days a week.
The inscriptions on the exterior of the meter or lancing
device have been removed.
The color of the meter or lancing device has changed or faded.
Cracks or roughness develop on the meter or lancing device.
A part of the segment on the meter display does not flash.
Control solution test results do not fall within the stated
range on the test strip vial.