Pre-cleaning and Disinfection Procedures
1
Open the cap of the Clorox® Germicidal
Wipes container and pull out one (1)
towelette and close the cap.
2
Wipe the entire surface
of the meter three (3)
times horizontally and
three (3) times vertically
using one towelette to
pre-clean blood and
other body fluids.
3
Dispose of the used towelette in a trash bin.
4
Pull out one (1) new towelette and
wipe the entire surface of the
meter three (3) times
horizontally and three (3)
times vertically using a new
towelette to disinfect
the meter.
5
Dispose of the used towelette in a trash bin.
6
Allow exteriors to remain wet for one (1)
minute, then wipe the meter using a dry cloth.
Front
Back
Both sides