60
Precision
Precision studies were performed in a laboratory using the
ReliOn™ Premier BLU Blood Glucose Monitoring System.
User performance results for glucose concentrations between
48 mg/dL and 553 mg/dL.
Within ± 15 mg/dL and Within ± 20%
371/371 (100%)
Within Run Precision
Blood avg. 43 mg/dL SD = 1.9 mg/dL CV = 4.3%
Blood avg. 71 mg/dL SD = 2.1 mg/dL CV = 2.9%
Blood avg. 135 mg/dL SD = 3.8 mg/dL CV = 2.8%
Blood avg. 203 mg/dL SD = 5.2 mg/dL CV = 2.6%
Blood avg. 343 mg/dL SD = 11 mg/dL CV = 3.2%
Between Run Precision
Control avg. 36 mg/dL SD = 1.4 mg/dL CV = 3.8%
Control avg. 114 mg/dL SD = 3.4 mg/dL CV = 3.0%
Control avg. 341 mg/dL SD = 8.2 mg/dL CV = 2.4%