Control Solution Testing
Insert a test strip into the meters test strip
port with the contact bars facing upwards.
Gently push the test strip into the port until
the meter beeps. Be careful not to bend the
test strip while pushing it in.
The
symbol will show up.
You want to practice the test procedure using the control
solution, instead of blood.
You use the meter for the first time.
You begin using a new vial of test strips.
You suspect that the meter or test strips are not
working properly.
You think your test results are inaccurate or they do not reflect
how you feel.
• The meter is dropped or damaged.
Use ReliOn Premier Control Solution when:
Step 1
You can flag the control solution test result by
pressing the > button for 3 seconds. To undo
the control solution flag, press the > button
for 3 seconds again.
Step 2