REF 1029568 Rev A BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc. 7-1
Chapter 7. Performance Verification
The performance verification verifies the integrity of the ventilator's critical
components using external measurement devices. The performance
verification consists of several tests (Table 7-1).
The type of service performed determines which of the tests are required
(Table 7-2). Run required tests in order.
Performance Verification Tests
Test number Description
1 Electical Safety
2Alarm / LED Test
3 BiPAP Focus System Run-In
4 Real Time Clock Calibration
5 BiPAP Focus Calibration
6 Performance Verification
Table 7-1: Performance Verification: Required Tests
Performance Verification: Required Tests
Service performed Required tests
Annual or two-year preventive
maintenance
All.
Removal/replacement
Controller PCB All.
User Interface PCB All.
Internal battery Alarm/keypad/LED, power transition.
VGA PCB Electical Safety, Alarm/keypad/LED, power
transition.
LCD assembly All.
keypad Electical Safety, Alarm/keypad/LED.
main alarm, backup alarm, or remote
alarm connector
Electical Safety, Alarm/keypad/LED, power
transition.
Table 7-2: Performance Verification: Required Tests