Chapter 7
Performance Verification
REF 1029568 Rev A BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc. 7-7
Figure 7-2: Begin Calibration
9. In the upper left-hand corner below the menu bar, click the white
run time arrow. The program will set the appropriate date and time on
the BiPAP Focus.
10. When the program has finished, a green "PASS" indicator will appear
at the bottom.
Figure 7-3: CALIBRATION COMPLETE