2 Command System RIGOL
MSO2000A/DS2000A Programming Guide
2-139
:MEASure:STATistic:DISPlay
Syntax
:MEASure:STATistic:DISPlay <bool>
:MEASure:STATistic:DISPlay?
Description
Enable or disable the statistic function; or query the current status of the statistic function.
Parameter
Name Type Range Default
<bool> Bool {{0|OFF}|{1|ON}} 0|OFF
Explanation
When the statistic function is enabled, the system will make statistics and display the measurement
results of at most five measurement items that are enabled last. When the statistic type is extrum, the
statistic results include the current value, average, minimum and maximum. When the statistic type is
difference, the statistic results include the current value, average, count and standard deviation.
Return Format
The query returns 0 or 1.
:MEASure:STATistic:MODE
Syntax
:MEASure:STATistic:MODE <mode>
:MEASure:STATistic:MODE?
Description
Set or query the statistic mode.
Parameter
Name Type Range Default
<mode> Discrete {DIFFerence|EXTRemum} EXTRemum
Explanation
EXTRemum: display the maximum and minimum values.
DIFFerence: display the standard deviation and count.
Before using thoes command, enable the statistic function (refer to the
:MEASure:STATistic:DISPlay
command).
Return Format
The query returns DIFF or EXTR.

Table of Contents

Other manuals for Rigol 2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol 2000 Series and is the answer not in the manual?

Rigol 2000 Series Specifications

General IconGeneral
BrandRigol
Model2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals