RIGOL 2 Command System
2-154 MSO2000A/DS2000A Programming Guide
:MEASure:FREQuency
Syntax 1
:MEASure:FREQuency <chan>
Description 1
Enable the frequency measurement function of the specified channel.
Syntax 2
:MEASure:FREQuency? <chan>
Description 2
Enable the frequency measurement function of the specified channel and return the measurement result
(the unit is Hz).
Parameter
Name Type Range Default
<chan> Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|
D12|D13|D14|D15|CHANnel1|CHANnel2|MATH}
--
Return Format
The query returns the measurement result in scientific notation.
Example
:MEASure:FREQuency CHANnel1
:MEASure:FREQuency? CHANnel1
You can use the following commands to query the statistic maximum, minimum, current value, average
and standard deviation of the frequency measurement values of the specified channel waveform. Sending
the following commands will enable the statistic function automatically.
:MEASure:FREQuency:SMAXimum?
:MEASure:FREQuency:SMINimum?
:MEASure:FREQuency:SCURrent?
:MEASure:FREQuency:SAVerage?
:MEASure:FREQuency:SDEViation?

Table of Contents

Other manuals for Rigol 2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol 2000 Series and is the answer not in the manual?

Rigol 2000 Series Specifications

General IconGeneral
BrandRigol
Model2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals