RIGOL 2 Command System
2-12 MSO2000A/DS2000A Programming Guide
:ACQuire:MDEPth
Syntax
:ACQuire:MDEPth <mdep>
:ACQuire:MDEPth?
Description
Set or query the memory depth of the oscilloscope; namely the number of waveform points that can be
stored in a single trigger sample. The default unit is pts (points).
Parameter
Name Type Range Default
<mdep> Discrete Refer to the Explanation AUTO
Explanation
When a single channel is on: AUTO|14000|140000|1400000|14000000|56000000 (option).
When dual channels are on: AUTO|7000|70000|700000|7000000|28000000 (option).
Return Format
The query returns the actual number of points (integer) or AUTO.
:ACQuire:SRATe?
Syntax
:ACQuire:SRATe?
Description
Query the current sample rate. The default unit is Sa/s.
Return Format
The query returns the current sample rate in scientific notation. For example, 1.00000e+09; namely