EasyManuals Logo

Rigol 2000 Series Programming Guide

Rigol 2000 Series
376 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #324 background imageLoading...
Page #324 background image
RIGOL 2 Command System
2-308 MSO2000A/DS2000A Programming Guide
:TRIGger:SLOPe:TLOWer
Syntax
:TRIGger:SLOPe:TLOWer <time>
:TRIGger:SLOPe:TLOWer?
Description
Set or query the lower limit of time in slope trigger. The unit is s.
Parameter
Name Type Range Default
<time> Real
or : 10ns to 1s
or : 10ns to 999ms
1us
Explanation
This command is only available when the trigger condition (refer to the
:TRIGger:SLOPe:WHEN command)
is PGReater, NGReater, PGLess or NGLess.
Return Format
The query returns the lower limit of time in scientific notation.
:TRIGger:SLOPe:TUPPer
Syntax
:TRIGger:SLOPe:TUPPer <time>
:TRIGger:SLOPe:TUPPer?
Description
Set or query the upper limit of time in slope trigger. The unit is s.
Parameter
Name Type Range Default
<time> Real
or : 10ns to 1s
or : 20ns to 1s
2us
Explanation
This command is only available when the trigger condition (refer to the
:TRIGger:SLOPe:WHEN command)
is PLESs, NLESs, PGLess or NGLess.
Return Format
The query returns the upper limit of time in scientific notation.

Table of Contents

Other manuals for Rigol 2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol 2000 Series and is the answer not in the manual?

Rigol 2000 Series Specifications

General IconGeneral
BrandRigol
Model2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals