RIGOL 2 Command System
2-22 MSO2000A/DS2000A Programming Guide
:BUS<n>:PARallel:THReshold
Syntax
:BUS<n>:PARallel:THReshold <sour>,<thre>
:BUS<n>:PARallel:THReshold? <sour>
Description
Set or query the analog channel threshold of parallel decoding on the specified bus.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<sour> Discrete {CHANnel1|CHANnel2} --
<thre> Real -5×VerticalScale–OFFSet to +5×VerticalScale–OFFSet 0
Note: For the VerticalScale, refer to the :CHANnel<n>:SCALe command. For the OFFSet, refer to
the
:CHANnel<n>:OFFSet command.
Return Format
The query returns the current threshold in scientific notation.