RIGOL
© 2008 RIGOL Technologies, Inc.
User‟s Guide for Logic Signal Output Module
2-43
Digital Waveform Calibration
Digital Waveform Calibration is used to calibrate the output voltage of the module.
When users connect the digital module with different DG3000s or connect a
DG3000 with different modules, the calibration should be performed.
Press Arb Cal Digital W, to enter the interface as shown in Figure 2-47.
Figure 2-47 Digital waveform calibration interface
Figure 2-48 Table 2-13 Operation menu description
Menu
Setting
Description
Calibrate
High volt
Select the High Level to be
calibrated.
Calibrate
Low volt
Select the Low Level to be
calibrated.
Set to
Default
Restore to default calibration.
YES
Press the button to complete the
selected calibration item.
End
Cal
Finish the calibration.