17. Trigger Source of Digital Logic Signal has been changed to Manual.
Press Burst Source“Internal”, press Trigger on the front panel of the
instrument; this message will appear.
18. Set the rate by bit per second (bps).
Press Arb Edit Digital WRS232Baud, choose the “User” option; this
message will appear.
19. High level has been calibrated.
Press Arb Cal Digital WCalibration High volt, input the measured value,
and press YES; this message will appear.
20. Low level has been calibrated.
Press Arb Cal Digital WCalibration Low volt, input the measured value,
and press YES; this message will appear.
21. Please measure A0 channel, then input the value.
Press Arb Cal Digital W Calibration High volt, (Calibration Low volt); this
message will appear.
22. Voltage calibration coefficients are restored to factory defaults.
Press Arb Cal Digital WSet to Default; this message will appear.
23. Please choose calibration item.
Press Arb Cal Digital WSet to DefaultYES; this message will appear.
24. No changes.
When you are changing certain parameter, and press CANCEL; this message
will appear to prompt users that there is not changed.
(**denotes the content will vary according to the setting)