RIGOL Chapter 5 IQ Modulation Output
User’s Guide for DG5000
5-6
User-defined Code Pattern
User could edit code pattern with up to 2000 Words (4000 Bytes) in the user space
provided by DG5000.
In internal modulation mode, press Code Pattern User to open user-defined code
pattern interface.
Figure 5-1 User-defined Code Pattern
As shown in the figure above, the data in Address Spacecorresponds to the data in
Data Spacerespectively. A hexadecimal data occupies two address spaces, for
example, the lower 8 bits of data “0008”, “08”, is stored in the address “00000000”;
and the higher 8 bits, 00, is stored in the address “00000001”.
Edit User’s Data
1. Set Data Length
As mentioned before, you could edit users data with up to 4000 Bytes length. If
you need to set the length of your data, please press Length and use the
numeric keyboard to enter your desired length.
Address Space Data Space Data Capacity Data Edit Area