Chapter 3 Arbitrary Waveform Output RIGOL
User’s Guide for DG5000
3-15
Edit Points
The following example introduces how to create an arbitrary waveform which meets
with the following conditions by using Edit Points”.
Operating Procedures:
1. Press Arb, use to open the 2/2 menu page, and then press Create New
Cycle Period, use the numeric keyboard to input “12” and select the unit μs
in the pop-up menu.
2. Press High VLimit, use the numeric keyboard to input “4” and select the unit “V”
in the pop-up menu.
3. Press Low VLimit, use the numeric keyboard to input “-2” and select the unit “V”
in the pop-up menu.
4. Press Init #Points, use the numeric keyboard to input4” and then press OK.
Note that a level line of -2 V appears.
5. Press Interp to selectLinear.
6. Press Edit Points to enter the Edit Pointinterface.
1) Press PointID to define the first point (default time is 0). Press Voltage and
Parameter Value
Cycle Period 12 μs
High VLimit 4 V
Low VLimit -2 V
Interpolation Linear
Point 1 0 s, 0 V
Point 2 4 μs, 4 V
Point 3 8 μs, 0 V
Point 4 10 μs, -2 V