Chapter 2 Basic Waveform Output RIGOL
User’s Guide for DG5000
2-9
Recover Delay
The recover delay function is only for dual-channel models. Press Pulse, then use
to open the 2/2 menu page and press Restore, the generator will align the
delay between the two channels. Assume that CH1 and CH2 output pulse waveforms
with the same parameters. Use the oscilloscope to sample the waveforms of the two
channels and display them stably. Then, toggle the output switch of the generator.
The two waveforms displayed on the oscilloscope have a certain delay. At this point,
press Restore on the generator and the waveforms will be displayed on the
oscilloscope without any delay.
Before Restore
After Restore
Note: For dual-channel models, the corresponding Restore is grayed out and
disabled if any of the two channels is in modulation mode.
CH1
CH2
CH1
CH2