119Chapter 6: Maintenance GX-3R Pro Operator’s Manual
Replacing the Batteries (Alkaline Version)
NOTE: Use Duracell MN2400 or PC2400 alkaline batteries or RKI Instruments, Inc. lithium-
ion battery pack 49-1625 to maintain the CSA classification of the GX-3R Pro. Use of
other batteries will void the CSA classification and may void the warranty. Do not mix
old/new or different types of batteries.
NOTE: Utiliser Procell 2400 piles alcalines ou RKI Instruments, Inc. pack batterie lithium-ion
49-1625 de maintenir la classification CSA de la GX-3R Pro. L’utilisation d’autres piles
annule la classification CSA et peut annuler la garantie. Ne mélangez pas les anciennes/
nouvelles ou différents types de piles.
CAUTION: Sensors, filters, and alkaline batteries are the only user-serviceable parts in the GX-
3R Pro. Do not replace or modify any other part.
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed in
an area known to be nonhazardous.
AVERTISSEMENT: Pour éviter l'inflammation d'une atmosphère dangereuse, les piles ne
doivent être remplacées que dans une zone non dangereuse.
Replace the 2 AAA batteries when the battery icon indicates that the unit is in low battery
warning. When in low battery warning, the lowest battery level indication bar disappears and the
battery icon will be flashing.
1. Turn off the GX-3R Pro.
WARNING: Do not remove the batteries while the GX-3R Pro is on.
2. Use a coin or a screwdriver to turn the battery cover counterclockwise.
3. Carefully remove the old alkaline batteries. Verify that the battery compartment and electrical
contacts are clean.