55Chapter 4: Display Mode GX-3R Pro Operator’s Manual
Alarm Points Screen (ALARM POINTS)
The Alarm Points Screen shows the gas alarm settings for all active channels.
1. While in Display Mode, press and release POWER MODE until ALARM POINTS appears.
2. Press and release AIR to enter the Alarm Points screen. The Full Scale Settings screen appears
and shows full scale settings for each channel.
3. Use AIR to scroll through the Warning, Alarm, Alarm H, STEL, and TWA settings.
7:49
YES : AIR
NO : MODE
ALARM POINTS
AIR
AIR
AIR
AIR
AIR
FULL SCALE
7:49
40.0 10.00
O2 % CO2 vol%
100 2000 200.0
CH4 %LEL CO ppm H2S ppm
WARNING
7:49
19.5 0.50
O2 % CO2 vol%
10 25 5.0
CH4 %LEL CO ppm H2S ppm
ALARM
7:49
18.0 3.00
O2 % CO2 vol%
25 50 30.0
CH4 %LEL CO ppm H2S ppm
ALARM H
7:49
23.5 3.00
O2 % CO2 vol%
50 1200 100.0
CH4 %LEL CO ppm H2S ppm
STEL
7:49
3.00
CO2 vol%
200 5.0
CO ppm H2S ppm
TWA
7:49
0.50
CO2 vol%
25 1.0
CO ppm H2S ppm