EasyManua.ls Logo

RKI Instruments GX-3R Pro - Updating the Lunch Break Setting (LUNCH BREAK); Setting the Confirmation Beep and Non-Compliance Indicator (CONFIRMATION)

RKI Instruments GX-3R Pro
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
105Chapter 5: User Mode and Calibration GX-3R Pro Operator’s Manual
Updating the Lunch Break Setting (LUNCH BREAK)
OFF (factory setting): The GX-3R Pro automatically starts new TWA and PEAK reading
collection and resets the time in operation at startup.
ON: The Lunch Break Screen displays during startup. From this screen, you can choose to
continue accumulating TWA and PEAK readings and the time in operation from the last time the
GX-3R Pro was used or start collecting new readings and reset the time in operation.
1. While in User Mode, use AIR to place the cursor next to LUNCH BREAK.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to User Mode.
5. See “Returning to Measuring Mode (START MEASURE)” on page 117 to return to
Measuring Mode.
Setting the Confirmation Beep and Non-Compliance
Indicator (CONFIRMATION)
The CONFIRMATION menu item has 3 items: BEEP SELECT, BEEP INTERVAL, and
ESCAPE.
1. While in User Mode, use AIR to place the cursor next to CONFIRMATION.
2. Press and release POWER MODE to enter the Confirmation Menu.
MAINTE 7:49
ALARM SETTING
LUNCH BREAK
CONFIRMATION
USER MODE
MAINTE 7:49
LUNCH BREAK
CONFIRMATION
AUTO BACKLIGHT
USER MODE
MAINTE 7:49
BEEP SELECT
BEEP INTERVAL
ESCAPE
CONFIRMATION

Table of Contents