EasyManua.ls Logo

RKI Instruments GX-3R Pro - User Mode Zero Follower (DISP ZERO FLWR); Setting the Date;Time (DATE)

RKI Instruments GX-3R Pro
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
181Appendix B: Maintenance Mode GX-3R Pro Operator’s Manual
User Mode Zero Follower (DISP ZERO FLWR)
ON: Zero follower menu item appears in User Mode.
OFF (factory setting): Zero follower menu item does not appear in User Mode. Zero follower
menu item is always available in Maintenance Mode.
It is not normally necessary to have the zero follower menu item appear in User Mode. Contact
RKI Instruments before turning this setting on.
Setting the Date/Time (DATE)
1. While in Maintenance Mode, use AIR to place the cursor next to DATE.
2. Press and release POWER MODE. The date and time appear and the year flashes.
3. Use AIR to display the desired year.
4. Press and release POWER MODE to save the setting. The month setting flashes.
5. Repeat Step 3 and Step 4 to enter the month, day, hours, and minutes settings. The date and
time are saved and the instrument returns to Maintenance Mode.
6. See “Exiting Maintenance Mode (START MEASURE)” on page 189 to return to Measuring
Mode.
MAINTE 7:49
DISP ZERO FLWR
DATE
DATE FORMAT
MAINTENANCE
MAINTE 7:49
MM/DD/YYYY
1/14/2019
7:49
DATE

Table of Contents