GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration92
4. Press and release POWER MODE to save the setting and return to the CAL SETTING menu.
5. See “ESCAPE” on page 93 to return to User Mode.
ESCAPE
1. While in the CAL SETTING menu, use AIR to place the cursor next to ESCAPE.
2. Press and release POWER MODE. The instrument returns to User Mode.
3. See “Returning to Measuring Mode (START MEASURE)” on page 117 to return to
Measuring Mode.
Setting Bump Test Parameters (BUMP SETTING)
The BUMP SETTING menu item has 6 items: BUMP PARAMETERS, BUMP REMINDER,
BUMP INTERVAL, BUMP EXPIRED, BUMP CHECK GAS, and ESCAPE.
1. While in User Mode, use AIR to place the cursor next to BUMP SETTING.
2. Press and release POWER MODE. The BUMP SETTING menu appears.
BUMP PARAMETER
The BUMP PARAMETER menu item has 5 items: GAS TIME, TOLERANCE, CAL TIME,
AUTO CAL, and ESCAPE.
1. While in the BUMP SETTING menu, use AIR to place the cursor next to BUMP
PARAMETER.
MAINTE 7:49
CAL SETTING
BUMP SETTING
MAN DOWN
USER MODE
MAINTE 7:49
BUMP PARAMETER
BUMP REMINDER
BUMP INTERVAL
BUMP SETTING
MAINTE 7:49
BUMP PARAMETER
BUMP REMINDER
BUMP INTERVAL
BUMP SETTING