GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration106
BEEP SELECT
BEEP SELECT defines what kind of confirmation or non-compliance indication you want to
occur in Measuring Mode. The available choices are:
OFF (factory setting): The GX-3R Pro does not provide a confirmation alert or non-compliance
indicator.
LED: The GX-3R Pro’s LEDs double flash as often as defined by the BEEP INTERVAL
parameter to verify that the instrument is operating.
BUZZER: The GX-3R Pro’s buzzer double beeps as often as defined by the BEEP INTERVAL
parameter to verify that the instrument is operating.
LED+BUZZER: The GX-3R Pro’s LEDs double flash and the buzzer double beeps as often as
defined by the BEEP INTERVAL parameter to verify that the instrument is operating.
BUMP/CAL: If a bump test or a calibration is due and if BUMP EXPIRED or CAL EXPIRED
is set to CONFIRM TO USE or NO EFFECT, the GX-3R Pro’s LEDs double flash as often as
defined by the BEEP INTERVAL parameter to indicate a non-compliance. Once a successful
bump test or calibration (depending on which is due) is done, the LEDs stop flashing.
ALARM ALERT
: If the instrument goes into any gas alarm, the LEDs double flash as often as
defined by the BEEP INTERVAL parameter to indicate a non-compliance. Once a successful
bump test or calibration is done, the LEDs stop flashing.
NOTE: Depending on the Need to get Bump Log setting (only accessed via SDM-3R
program), a data download may also be needed to clear the non-compliance.
BUMP/CAL/ALARM: The LEDs double flash to indicate a non-compliance if any of the
following happens.
a. BUMP EXPIRED is set to CONFIRM TO USE or NO EFFECT and a bump test is
due (cleared by successful bump test).
b. CAL EXPIRED is set to CONFIRM TO USE or NO EFFECT and a calibration is due
(cleared by successful calibration).
c. The instrument goes into any gas alarm (cleared by successful bump test or calibration).
NOTE: Depending on the Need to get Bump Log setting (only accessed via SDM-3R
program), a data download may also be needed to clear the gas alarm non-compliance.
1. While in the CONFIRMATION menu, use AIR to place the cursor next to BEEP SELECT.
2. Press and release POWER MODE. The current setting flashes.
MAINTE 7:49
BEEP SELECT
BEEP INTERVAL
ESCAPE
CONFIRMATION