GX-3R Pro Operator’s Manual Chapter 3: Operation24
4. If BUMP REMINDER is set to ON (factory setting is OFF) and a bump test is due, the
screen that appears next depends on how BUMP EXPIRED is set in User Mode (see page
97). The three possible screens are described below. If a bump test is not due, the instrument
displays the number of days left until a bump test is due.
BUMP EXPIRED set to
CONFIRM TO USE (factory set-
ting)
BUMP EXPIRED set to
CANNOT USE
BUMP EXPIRED set to
NO EFFECT
LCD
Sound Buzzer sounds double pulsing tone Buzzer sounds double pulsing tone None
Action Option A, Perform bump test
:
Press and release POWER MODE
to perform a bump test. The
instrument takes you straight to the
BUMP TEST CYLINDER A
screen in the BUMP TEST User
Mode item. If Password
Protection is set to On using the
SDM-3R PC program, you must
enter a password. See page 68 for
bump test instructions.
If the bump test is successful, the
screen above will not appear again
until the unit is due for bump
testing. If the bump test is not
successful, the screen above will
again appear in the startup
sequence.
Option B, Bypass message
: To
continue without performing a
bump test, press and release AIR.
The GX-3R Pro cannot be used until
a successful bump test is performed.
Press and release POWER MODE to
perform a bump test. The instrument
takes you straight to the BUMP
TEST CYLINDER A screen in the
BUMP TEST User Mode item. If
Password Protection is set to On
using the SDM-3R PC program, you
must enter a password. If you don’t
press POWER MODE, the instru-
ment automatically goes to the
BUMP TEST CYLINDER A screen
after 6 seconds. If Password Protec-
tion is set to On using the SDM-3R
PC program, you must enter a pass-
word. See page 68 for bump test
instructions.
If the bump test is successful, the
screen above will not appear again
until the unit is due for bump testing.
If the bump test is not successful, the
screen above will again appear in the
startup sequence.
Option A, Perform bump test
: To
perform a bump test, press and
release POWER MODE. The
instrument takes you straight to the
BUMP TEST CYLINDER A
screen in the BUMP TEST User
Mode item. If Password
Protection is set to On using the
SDM-3R PC program, you must
enter a password.
Option B, Bypass message
: To
continue without performing a
bump test, wait a few seconds for
the instrument to continue with its
startup sequence.

Table of Contents