GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration70
12. At the end of the countdown, the instrument analyzes the results. Follow the flow chart to
determine the bump test outcome.
Figure 8: Bump Test Flow Chart
Bump test passed?
Y N
AUTO CAL On?
Multiple Cylinders
Calibration passed?
Y N Y N
Assigned?
(default is on)
Multiple Cylinders
Assigned?
Y N
Y N
set to on? (default is on)
A D
F
B C E
G H
Y N Y N
Auto start after successful
bump
set to on? (default is on)
*NOTE: Auto start after successful bump/cal parameters set using Datalogging or SDM-3R PC Programs.
Auto start after successful
cal