73Chapter 5: User Mode and Calibration GX-3R Pro Operator’s Manual
Option G from Flow Chart Option H from Flow Chart
•Bump test failed
A-CAL set to ON (factory setting)
Calibration passed
One cylinder assigned
Auto Start After Successful Cal set to ON
(factory setting)
Bump test failed
A-CAL set to ON (factory setting)
Calibration passed
One cylinder assigned
Auto Start After Successful Cal set to
OFF (factory setting is ON)
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the GAS TIME and the CAL
TIME values defined in the BUMP
SET\SETTINGS item in User Mode.
3. The instrument shows which channels
passed or failed the bump test/
calibration. Use AIR to scroll between
the results and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
enter Measuring Mode.
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the GAS TIME and the CAL
TIME values defined in the BUMP
SET\SETTINGS item in User Mode.
3. The instrument shows which channels
passed or failed the bump test/
calibration. Use AIR to scroll between
the results and the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the START item in the BUMP
menu.
8. Press and release POWER MODE to
enter Measuring Mode.
MAINTE 60
30 2.0
CH4 %LEL H2S ppm
AUTO CAL APPLY GAS
AIR
AIR
MAINTE 7:49
P
O2 %
F P P F P
CH4 %LEL CO ppm H2S ppm
RESULT
MAINTE 7:49
11.9
O2 %
30 47 12.0
CH4 %LEL CO ppm H2S ppm
BUMP TEST
MAINTE 7:49
47 20.5
CH4 %LEL H2S ppm
AUTO CAL
MAINTE 60
30 2.0
CH4 %LEL H2S ppm
AUTO CAL APPLY GAS
AIR
AIR
MAINTE 7:49
P
O2 %
F P P F P
CH4 %LEL CO ppm H2S ppm
RESULT
MAINTE 7:49
11.9
O2 %
30 47 12.0
CH4 %LEL CO ppm H2S ppm
BUMP TEST
MAINTE 7:49
47 20.5
CH4 %LEL H2S ppm
AUTO CAL