GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration • 84
Setting the Calibration Values in SETTING CAL--P
1. While in User Mode, use AIR to place the cursor next to GAS CAL.
2. Press and release POWER MODE. The Gas Cal menu appears.
3. Use AIR to place the cursor next to AUTO CAL.
4. Press and release POWER MODE. The Auto Cal screen for gases assigned to Cylinder A
displays.
5. Use AIR to scroll to SETTING CAL--P.
6. Press and release POWER MODE. The combustible gas channel displays.
MAINTE 7:49
BUMP TEST
GAS CAL
CAL SETTING
USER MODE
MAINTE 7:49
CO2 ZERO CAL
AUTO CAL
ESCAPE
GAS CAL
MAINTE 7:49
SETTING CAL--P
MAINTE
- - - -
CH4 %LEL
SETTING CAL--P