GX-3R Pro Operator’s Manual Chapter 5: User Mode and Calibration • 90
CAL REMINDER
ON (factory setting): The GX-3R Pro gives an indication at start up if it is due for calibration. The
type of indication depends on the CAL EXPIRED setting (see page 91).
OFF: The GX-3R Pro does not give an indication at start up if it is due for calibration.
1. While in the CAL SETTING menu, use AIR to place the cursor next to CAL REMINDER.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL SETTING menu.
5. See “ESCAPE” on page 93 to return to User Mode.
CAL INTERVAL
This setting defines the amount of time between calibrations. The time can be set in 1 day
increments. The minimum setting is 1 day and the maximum setting is 1000 days. The factory
setting is 90 days.
1. While in the CAL SETTING menu, use AIR to place the cursor next to CAL INTERVAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL SETTING menu.
5. See “ESCAPE” on page 93 to return to User Mode.