EasyManua.ls Logo

Rosemount 3095 - Page 63

Rosemount 3095
115 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-7
Troubleshooting and Maintenance
TABLE 5-5. (continued).
Symptom Corrective Action
Erratic PV Reading LOOP WIRING
CAUTION
Do not use higher than the specified voltage to check the loop
or damage to the device may result.
Check for adequate voltage to the device. It should be
11 to 55 V dc at the device terminals.
• Check for intermittent shorts, open circuits, and
multiple grounds.
PROCESS PULSATION
Adjust the electronic damping (see page 4-4).
DEVICE ELECTRONICS
Connect a Model 275 to check the sensor limits to ensure
calibration adjustments are within the sensor range and that
calibration is correct.
Make sure the post connectors are clean.
Confirm that the electronics housing is properly sealed
against moisture.
If the electronics are still suspect, substitute new electronics.
IMPULSE PIPING
Check for entrapped gas in liquid lines and for liquid in
gas lines.
Make sure that process fluid has not frozen within the
process flange.
SENSING ELEMENT
The sensing element is not field repairable and must be
replaced if found to be defective. Check for obvious defects,
such as a punctured isolating diaphragm or fill fluid loss, and
contact your nearest Rosemount Service Center.

Table of Contents

Related product manuals