EasyManua.ls Logo

Rosemount Oxymitter 5000 - Manual Calibration with Membrane Keypad

Rosemount Oxymitter 5000
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hazardous Area Oxymitter 5000
9-4
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
3. IMPS 4000. Use the IMPS 4000 keypad to change the InitCalX
parameter of the CHANGE PRESETS display mode from 0000 to 0001.
Refer to the IMPS 4000 Intelligent Multiprobe Test Gas Sequencer
Instruction Bulletin for more information.
4. FOUNDATION fieldbus. Use fieldbus to perform the O
2
CAL method.
5. Remote Contact. Initiate a calibration from a remote location via the
remote contact input connection provided by an IMPS 4000 or SPS
4001B. Refer to the documentation available for the control system in
use for more information.
Once a semi-automatic calibration is initiated by any of the methods
previously described, the Hazardous Area Oxymitter 5000's CALIBRATION
RECOMMENDED alarm signals an IMPS 4000 or SPS 4000 to initiate a
calibration. The sequencer sends an "in cal" signal to the control room so that
any automatic control loops can be placed in manual. Then, the sequencer
begins to sequence the calibration gases.
Manual Calibration with
Membrane Keypad
Manual calibrations must be performed at the Hazardous Area Oxymitter
5000 site and will require operator intervention throughout the process.
Manual calibration instructions, in condensed form, can also be found on the
inside of the right electronics housing cover (Figure 9-2).
Use the following procedure to perform a manual calibration:
1. Place control loop in manual.
2. Verify the calibration gas parameters are correct per "Calibration with
Keypad".
Figure 9-2. Inside Right Cover
SW2DIP SWITCH
LED
HEATER
O2CELL
CALIBRATION
HEATER T/C
STATUS
OPEN
SHORTED
REVERSED
HIGHHIGH
TEMP
HIGHCASE
TEMP
LOW TEMP
HIGH TEMP
OPEN
BAD
EPROM
CORRUPT
INVALIDSLOPE
INVALID
CONSTANT
FLASHES
1
2
3
4
1
2
OPEN
3
4
5
1
3
4
1
2
PUSHCAL
CAL LEDON
APPLY TG1
PUSHCAL
CAL LEDONSOLID
REMOVE TG1& APPLY TG2
PUSHCAL
CAL LEDONSOLID
WAIT FORFLASH
REMOVE TG2
PUSHCAL
1
2
3
4
5
6
7
ALARMS
MANUAL
CALIBRATION
CAL LEDONFOR
PURGE TIME
WAIT FORFLASH
CAL LEDOFF
PUSHCAL
CAL LEDFLASH
2FLASH-VALIDCAL
3FLASH-INVALIDCAL
8
PLACECONTROL LOOP
INMANUAL
IFCAL LEDON
GO TOSTEP 2
*
*
3
LAST CAL
FAILED
A/DCOMM
ERROR
38740018
OFF
ON
NOT USEDNOT USED
NOT USEDNOT USED
NOT USEDNOT USED

Table of Contents

Related product manuals