1.11 GENERAL
DOC01-V6ULB+ 11
1.11 Electromagnetic compatibility (EMC)
This hospital bed complies with the following electromagnetic compatibility standards.
Standards and test name Compliance
CISPR 11 (2015) A1 (2016) Conducted emissions Group 1 - Class B 150kHz-30MHz
CISPR11 (2015) A1 (2016) Radiated emissions (Prescan 3m) Group 1 - Class B 30MHz-1GHz
CISPR11 (2015) A1 (2016) Radiated emissions (10m) Group 1 - Class B 30MHz-1GHz
IEC 61000-3-2 (2018) Harmonic current emission limits Class A
IEC 61000-3-3 (2013) A1 (2017) Voltage fluctuations and flicker
limitations
Observation Pst: 10 min.
Observation Plt: 120 min
IEC 61000-4-2 (2008) Immunity to electrostatic discharge
Contact: ±8kV. Air: ±2kV, ±4kV, ±8kV, ±15kV
IEC 61000-4-3 (2020) Immunity to radiated electromagnetic fields
80MHz-2.7GHz: 3V/m
IEC 60601-1-2(2014) Proximity fields of RF equipment
385MHz: 27V/m
710MHz,745MHz, 780MHz:9V/m
450MHz, 810MHz, 870MHz, 930MHz: 28V/m
1.72GHz, 1.845GHz, 1.97GHz, 2.45GHz:
28V/m
5.24GHz, 5.5GHz, 5.785GHz: 9V/m
IEC 61000-4-4 (2012) Immunity to fast electrical transients
Power: ±2kV / 100kHz, I/O Ports: ±1kV /
100kHz, Communication Ports: N/A
IEC 61000-4-5 (2014) A1 (2017) Overvoltage immunity
Power: ±2kV L-PE / ±1kV LL, I/O Ports: N/A,
Communication Ports: N/A
IEC 61000-4-6 (2013) Immunity to conducted disturbances induced
by RF fields
Power: 3V, I/O Ports: 3V,
Communication Ports: 3V, ISM Bands: 6V
IEC 61000-4-8 (2009) Immunity to power-supply magnetic fields Continuous field: 30A/m / 50Hz & 60Hz
IEC 61000-4-11 (2020) Immunity to voltage dips, short
interruptions and voltage variations on AC input
Voltage drops:
0%One for half a cycle
0%One for 1 cycle
70%One for 25 cycles (at 50Hz)
Short interruption:
0%Un for 250 cycles (at 50Hz)
ELECTROMAGNETIC COMPATIBILITY WARNING
Portable RF communications equipment, including peripherals such as antenna
cables and external antennas, must not be closer than 30 cm (12 inches) to any
part of the bed, including cables.
Avoid stacking or placing equipment adjacent to other equipment to prevent
malfunctioning. If such use is necessary, observe stacked or adjacent equipment
carefully to ensure that it is operating correctly.
The use of accessories, transducers and cables, other than those specified or
supplied by the manufacturer, could result in increased electromagnetic
emissions or reduced electromagnetic immunity, and lead to incorrect operation.