3.14 OPERATING INSTRUCTIONS
30 DOC01-V6ULB+
3.14 Nurse call and DB37 output
Pressing the nurse call button sends a signal to the on-call station.
The bed must be connected to the wall using a DB37 communication wire. The socket
for this wire is located under the frame at the head end of the bed.
This type of connection is an industry standard. If another type of connection is required,
please contact us to find out about other available connection options.
Depending on the facility's system configuration, the socket offers the option of sending
a normally connected (NC) or normally open (NO) signal.
Note: This socket also sends a signal on bed exit if this option is enabled.
WARNINGS ON THE DB37 SOCKET
Check that the signal is working properly after connecting. A non-functional
connection could result IN PROCESSING DELAYS THAT COULD
AGGRAVATE OR CAUSE INJURY.