EasyManua.ls Logo

Rotec VersaTech 600 ULB - Preventive Check

Rotec VersaTech 600 ULB
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2 MAINTENANCE
DOC01-V6ULB+ 37
4.2 Preventive check
Annual checks:
Check that the unit is still in good condition;
Connect the device cable;
Leave the unit plugged in for at least 48 consecutive hours;
Check that bolts and nuts are tight;
Check shoulder washers used to reduce friction on moving parts for wear. Do not
lubricate. Replace if necessary;
Check that no objects or equipment hinder the movement of the bed's moving parts
(wheels, frame, etc.);
Check condition of remote control wire, power cord and bed wiring. Replace
damaged wires;
Check that all functions are operating correctly, in accordance with the operating
instructions in this manual (see section 3).
Disconnect the power cable and operate the motors to check the battery.
For optimum performance, we recommend replacing the bed's batteries every 3 years.
PREVENTIVE INSPECTION WARNINGS
Periodic maintenance is essential to maintain the bed's reliability. Any bed
showing signs of malfunction must only be used and repaired by authorized and
competent personnel. Failure to comply with these safety instructions could
jeopardize the integrity of the bed and result in SERIOUS INJURY.
Any replacement of equivalent parts not certified by Rotec International may
result in SERIOUS INJURY AND/OR SIGNIFICANT MECHANICAL BREAKAGE AND
MAY LIMIT OR VOID THE PRODUCT WARRANTY.
Any breakage of a part giving access to electronics must result in immediate
storage of the unit until the damaged part has been replaced. Continued use
under these conditions may result in SERIOUS ELECTRICAL INJURY.

Table of Contents