4.4 MAINTENANCE
DOC01-V6ULB+ 39
4.4 Troubleshooting guide
If you experience any difficulties with the bed and/or any of its components, please
consult the section below. If the following or other problems persist, do not hesitate to
contact our technical support department.
Checks
PROBLEMS VERIFICATIONS
None of the controls work, whether
Is the power cord properly connected?
Either on the remote control or on the
nursing control.
Is the power cord damaged?
If so, replace it.
Check all connection plugs.
No commands work on
Are the controls locked?
the remote control.
Is the remote control cable properly
connected?
Is the remote control cable damaged?
If so, replace it.
One or more functions of the
Check the above points.
remote control does not work.
Are the controls locked?
If no, on the nurse control commands
do they work?
If so, have the remote control checked.
The bed move slowly.
Is the power cord properly connected?
Is the power cord damaged?
If so, replace it.
The device emits one or more beeps.
See the following section for an
explanation of acoustic signals