1.2 GENERAL
6 DOC01-V6ULB+
1.2 Intended use
This manual has been designed to assist you in using the VersaTech 600 ULB+ bed from
Rotec International. Please read this document carefully before use to ensure safe and
secure operation.
This manual is an integral part of the unit and should always be included with the unit
when sold or transferred. It must be accessible at all times to medical and maintenance
personnel.
Area: This device has been designed for use in a care environment such as a hospital or
other medical facility, where medical supervision is required, where monitoring is
provided if necessary, and where the bed is used in procedures of a medical nature for
treatment, diagnosis, monitoring, helping to maintain or improve the patient's condition.
This includes intensive care, ambulatory care, and regular short- and long-term care. It
is not designed for home care. Under no circumstances should this device be used in the
presence of flammable anesthetic gases mixed with air or oxygen, or nitrous oxide.
Applicable parts: the parts of the device that come into contact with the patient and
operator are intended to be the headboard and footboard, the side rails, the interface,
the bed base and the accessories intended for use with the device.
Patient: It is expected that the patient will be able to use the unit's functions to a limited
extent from the internal controls. The patient is expected to be of adult age, over 4ft
(1.22M) in height and between 70 lb (32kg) and 500 lb (272kg) in weight. It is not
designed for patients with behavioral or mental health problems.
Operator: It is expected that the operator using this unit will be a healthcare professional
such as a nurse, doctor, attendant, etc. This operator must be able to understand and