2. Description of the Piranha
Hardware and Specifications
2014-06/5.5C
15
Piranha & QABrowser Reference Manual
Reference conditions
Temperature
+18 °C to +23 °C
Relative humidity
50 %
Air pressure
101.3 kPa
X-ray field size
Inside the Piranha top panel.
Calibration is done with field size typically 5 mm less than the size of the
top panel.
Radiation quality
Radiography
Mammography
CT
70 kV, 2.5 mm Al
28 kV, 30 µm Mo
120 kV, 2.5 mm Al
Note: The reference conditions are given in reference to the IEC61674 standard.
Physical dimensions
Detector area
3 × 21.1 mm
Detector position
10 mm below top panel, as indicated in figure below and by a 3 mm rim
on 3 edges.
Size
133 × 75 × 26 mm (5.2" × 2.9" × 1.02")
Weight
Approximately 405 g