EasyManuals Logo

RTI Piranha Reference Manual

RTI Piranha
181 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
4. Measurement Principles & Theory
Update Modes
2014-06/5.5C
77
Piranha & QABrowser Reference Manual
During the measurement the time left will
count down and it can be stopped by tapping
the Cancel button.
The measured time will sometimes not be
exactly what was set under settings, but the
measured value is the one used for all
calculations. The dose rate shown for Timed
mode will always be mean dose rate during
the measurement cycle, i.e. the measured
dose divided by the measured time.
The Timed mode can be very useful both for very low dose rate measurements as well
as for long duration measurements.
For extreme low-level dose measurements you can improve your reading by
subtracting the background level. First do a Timed measurement without exposing
the detector to radiation and then do the same with radiation. The timed mode will
use the same measuring time and the first reading can be subtracted from the first.
Just make sure not to do a Reset between these measurements, as the Reset will
also make an offset adjustment. Note also that low-level readings may give
inaccurate kV readings.
For long duration measurements, cases with slowly rising and falling output, or cases
with very low pulse rate, timed mode may also be useful. For instance on CT
machines where the rotation cannot be stopped.
4.3.2 Using Free Run Update Mode
Free run update mode works almost exactly as the ordinary Continuous update mode.
There are however two differences:
1. Since there is no trig level, you will be able to measure lower, but there will be no time
reading unless the signal goes over the trig level.
2. You can select a moving average function. This lets you set a time for moving
average, this time acts as a averaging window, moving through time.
Moving average
This function is intended for low level dose rate measurements where increased
sensitivity and stability is needed. The function uses a moving average algorithm where
the number of seconds is selected by the user.
During the reset process the user must make sure that the detector is not exposed to

Table of Contents

Other manuals for RTI Piranha

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RTI Piranha and is the answer not in the manual?

RTI Piranha Specifications

General IconGeneral
BrandRTI
ModelPiranha
CategoryMultimeter
LanguageEnglish