2. Description of the Piranha
Maintenance
2014-06/5.5C
35
Piranha & QABrowser Reference Manual
2.6.2 Managing Detector Calibrations
RTI Detector Manager is a special Windows software that gives an overview of all
calibrations for the detectors and probes in your system. You will find the RTI Detector
Manager on your Product CD, in the folder \Software\RTI Detector Manager\.,Start the
file RTI Detector Manager.exe by double-clicking it.
Select the instrument of interest (Piranha)
and click OK.
If no instrument appears, check the
communication cable and that the Piranha is
powered on, then click Rescan.
Next, the available detectors
are shown to the left. The
Internal detector is always
available, but external probes
will only show up if they are
attached.
When clicking a detector, the
available calibrations will
show up to the right (In this
case the Internal detector is
highlighted). The TV and TF
columns show an × if there
are calibrations for Tube
Voltage and/or Total
Filtration. The factor column
shows the calibration factor
(for dose in this case).
Here is another example
(Piranha Dose Probe). This
type of detector only contains
a calibration factor for dose.