2. Description of the Piranha
Hardware and Specifications
2014-06/5.5C
17
Piranha & QABrowser Reference Manual
Note 1: This is valid for a tube with 14° anode angle. The HVL for a 22° anode is typically 0,5 mm
lower (@ 80 kV, 3 mm TF).
Note 2: All kerma and kerma rate ranges, inaccuracy, and resolution figures are valid for product
version 2 and higher of the Piranha.
Note 3: The Kerma rate is calculated as the Kerma (Dose) divided by the Irradiation time. See
also Waveforms and Triggers .
Note 4: The HVL range is valid if also the TF is within its specified range. For high TF at high kV
the HVL range may be limited by this.
Mammography
Note 1: All kerma and kerma rate ranges, inaccuracy, and resolution figures are valid for product
version 2 and higher of the Piranha.
Note 2: The Kerma rate is calculated as the Kerma (Dose) divided by the Irradiation time. See
also Waveforms and Triggers .
Note 3: The M6 (W / 50 µm Rh), M10 (W / 50 µm Ag), and M15 (W / 0.70 mm Al) calibrations are
suitable for the Hologic Selenia Dimensions and Fuji Amulet.
Note 4: The M16 (W / 50 µm Ag (Sel)) and M17 (W / 50 µm Rh (Sel)) calibrations are suitable for
Hologic Selenia with W anode..
Note 5: The M7 (W / 0.5 mmAl) calibration is suitable for Philips MicroDose Mammography
(Sectra).