8 - 2 WS80A Service Manual
Caution
This chapter describes the assembly/disassembly process of WS80A. Refer to this chapter for
upgrade or repair of the hardware.
WARNING
■ Samsung Medison Global Technology Support Group or its authorized
engineers are allowed to repair or replace the parts of the products.
■ The interior of this product contains dangerous high-voltage electricity.
Do not disassemble the product; there is a risk of electrocution.
■ Do not wear an antistatic wrist strap while you are working when the product
is powered on; you may sustain an electrical injury.
Preparation
You will need the following items: A crosshead (+) screwdriver, antistatic gloves, and an
antistatic wrist strap.
Turn off the power of the product, and disconnect the battery
NOTE
Wear antistatic gloves and a wrist strap when you are disassembling or assembling
the product. They help prevent accidents that could injure the engineer, and
product failure caused by static electricity.
[Figure 8.1] Antistatic Gloves and Wrist Strap