Chapter 3. Installing the Product 3 - 5
Installing the Product
Installation Safety
In a hospital environment, dangerous electrical current may occur as a result of the potential
difference between a contactable conductive part and connected equipment in treatment
rooms. The solution to the problem is consistent equipotential bonding. Equipotential
terminals on medical equipment should be connected to the equipotential bonding network
in medical rooms as shown in the picture.
[Figure 3.3 Equipotential bonding]
CAUTION
If the product needs to be transported or stored for an
extended duration, the temperature and humidity of the
environment must be checked.
A sudden change in temperature may cause condensation
and lead to product failure.
Refer to "[Table 3-2. Operational Temperature of Product]" before turning the product on
.
Temperature -20 -15 -10 -5 0 5 10 ~ 35 45 50 55 60
Waiting time 16 10 8 6 4 2
Use
immediately
2 4 6 10
[Table 3-2. Operational Temperature of Product]

Table of Contents